Sample test volume 100μL

Aspirated Sample volume 150μL

Sample capacity 20 sample carousel

Test time Less than 3 minutes

Sample throughput can process 20 samples in less than 1 hour

Unit mOsm/kg H2O

Calibration low: 0 to 2000 mOsm/kg H2O

full: 0 to 4000 mOsm/kg H2O

Resolution1 mOsm/kg H2O

Calibrated 3-point Calibrated for low range

4-point for full range

Communication Onbord printer, 10/100Mbps Ethernet, 4 USB 1.0/1.1/2.0 integrated bar code scanner

Accuracy 0 to 400 mOsm/kg H2O: mean value ≤ 2 mOsm/kg H2O from nominal value 400-4000 mOsm: mean value ≤ 0.5% from normal value

Precision (Within-run) standard deviation ≤ 2 mOsm/kg H2O between 0 and 400 mOsm; standard deviation ≤ 0.5% of value between 400 to 4000 mOsm

Drift Less than 1 mOsm/kg H2O per month

Temperature effects1 Recalibration recommended for ambient temperature change 5°C (9°F) or greater

Storage temperature -40°C to +45°C (-40°F to +113°F)

Simple Viscosity Up to 20 mPa-s(20 cP) higher viscosities may affect pipettor accuracy

Pipettor Accuracy< 1%Electrical voltage100-240 VAC (50/60 Hz)

Power consumption 375W

Dimensions (D x W x H)20.5″ x 23.6″ x 22.8″ (52 cm x 60 cm x 58 cm)

Net weight 68 lb (31 kg)

Shipping weight 133 lb (60 kg)

 