0 item(s)
View cart
You have no items in your shopping cart.
Mettler Toledo XP2U Analytical Balance
Model | XP2U 11122430 |
---|---|
Capacity x Readability | 2.1 g x 0.0001 mg |
Repeatability at nominal load | 0.00025 mg |
Repeatability at low load (measured at) | 0.0002 mg (0.2 g) |
Linearity | 0.001 mg |
Eccentric load deviation (test load)1) | 0.0025 mg (1 g) |
Sensitivity offset | 1.5 x 10–5 |
Sensitivity temperature drift2) | 0.0001%/°C |
Sensitivity stability3) | 0.0001%/a |
Typical values4) | |
Repeatability (sd) | 0.00015 mg+2.5 x (10–8)·R_gr |
Differential linearity deviation (sd) | 8 x (10-14) gER_nt |
Differential eccentric load deviation (sd) | 8 x (10–7)·R_nt |
Sensitivity offset (sd)2) | 3 x (10–6)·R_nt |
Minimum weight* (according to USP) | 0.45 mg+7.5 x (10–5)·R_gr |
Minimum weight* (@ U = 1%, 2 sd) | 0.03 mg+5 x (10–6)·R_gr |
Settling time | < 10 s |
pan size Ø | 16 mm |