You have no items in your shopping cart.
The AUTOPOL V Automatic Polarimeter is designed for today’s FDA regulated analytical laboratories. Whether you are a Pharmaceutical, Nutraceutical or University Lab, the AUTOPOL V has the standard measurement and compliance features you must have, including:
- 21CFR Part 11 Compliant
- Temptrol™ Electronic Heating and Cooling
- Six standard wavelengths
- Touch panel user interface
- Designed for use in FDA regulated Pharmaceutical, Nutraceutical and cGMP/GLP laboratories
AUTOPOL V SpecificationsMeasuring ModeOptical Rotation
Specific Rotation
Specific Rotation Plus Concentration
User Defined Scale
Sugar Degrees
°Z(ISS)
Measuring ScaleDegrees Arc, % Concentration
Measuring Range± 89.9° Arc Optical Rotation
± 999.99° Arc Specific Rotation
0-99.9% Concentration
Resolution0.001° Arc Optical Rotation
0.001% Concentration
0.001 Specific Rotation
Reproducibility0.002° Arc
Standard Model Accuracy0.002° up to 1°, 0.2% up to 5°, 0.01° above 5°
AP Model SpecsAccuracy (546nm and 589nm): ±0.002° Arc over ±89.9° Arc
Accuracy for other wavelengths is the same as the standard model
Selectable Resolution 0.01°, 0.001°, 0.0001° Arc Optical Rotation
PrismGlan Thomson calcite quartz
Optical Wavelengths365nm, 405nm, 436nm, 546nm, 589nm, 633nm
(other wavelengths available)
Wavelength SelectionTouchscreen Selectable: 6 standard
TempTrol™ RangeAutomatic Electronic Heating & Cooling 15°-35°C (Standard) ±0.2° C (Standard)
TempTrol™ Accuracy±0.2°C
Temp. Probe Range10°-100°C
Temp. Probe Accuracy±0.1°C
Acid ResistanceHastelloy™ measurement cell and Silco Steel™
sample chamber (optional)
Measurement Time4°/sec slewing rate & 5 sec nominal settling time
Light SourceTungsten-halogen 6V, 20W, avg. 2,000 hr life
Sample ChamberAccepts sample tubes up to 200mm
Data Storage32 GB Non-removable Compact Flast
Communication Interface3 - USB Ports
2 - RS232 Ports
Ethernet Port for Network Connection
CalibrationAutomatic calibration by push-button
Operating SystemWindows® Embedded
Display8 inch color LCD with LED backlight providing 800 x 600 resolution with 400 nits of luminance
User InterfaceTouchscreen
Automatic Sensitivity ControlMeasures samples with transmittance as low as 0.01 % (up to O.D. 4.0)
Input Power100-240 VAC, 50/60 Hz
Operating Dimensions30"W x 17"H x 10.5"D,
762mm W x 432mm H x 267mm D
Shipping Dimensions30"W x 11.5"H x 17"D,
762mm W x 431.8mm H x 279.4mm D
Operating Weight85 lbs (39 kg)